physiology, pah

Cardiac Output

Cardiac Output Measurement
By Donald R. Elton, MD, FCCP
Lexington Pulmonary and Critical Care


Introduction
Cardiac output is the amount of blood ejected from the left ventricle in one minute and is measured in liters per minute. Under normal circumstances, the outputs of the left and right ventricles must be equal in the absence of abnormal shunts between the pulmonary and systemic circulatory systems.

Physiology
Along with left ventricular filling pressures (pulmonary capillary wedge pressure), the cardiac output is one of the few hemodynamic parameters that with today's technology requires the placement of a pulmonary artery catheter. Cardiac output is the product of heart rate and stroke volume. Heart rate is determined by both intrinsic pacemaker function and modulation by the autonomic nervous system. Stroke volume is dependent upon the degree of diastolic ventricular filling coupled with the degree of contraction sometimes expressed as ejection fraction. Disease states can alter all of these components of cardiac output. Normally, as heart rate increases, the cardiac output increases proportionately. As heart rate increases however, the time available for ventricular filling to occur decreases and in each patient, there is a heart rate, above which, ventricular filling will decrease enough that further increases in heart rate will result in a lowered cardiac output. In a normal person, this cut-off occurs somewhere between 180-200 beats per minute while in disease states such as congestive heart failure secondary to cardiomyopathy, this cut-off may be reached at rates as low as 120 beats per minute. Uncontrolled atrial fibrillation or atrial flutter frequently result in heart rates that are too high for adequate cardiac output and a major part of the treatment of these arrythmias is to give the patient digoxin to help slow the abnormally high heart rate.

Measurement Methods

Two main methods are used to measure cardiac output today. These are the Fick method and dilution methods (either dye or thermal).

Fick Method

The Fick method requires that you be able to measure the A-V oxygen content difference and requires that you be able to measure the oxygen consumption. An arterial blood gas from a peripheral artery provides the blood for the CaO2 measurement or calculation while blood from the distal PA port of a Swan-Ganz catheter provides the blood for the CvO2 measurement or calculation. Oxygen consumption is obtained by measuring the inspired oxygen concentration and the expired oxygen concentration along with the expired minute volume. Small errors in the oxygen concentration measurements can result in large mathematical errors therefore these measurements should be made with a calibrated blood gas machine equipped for measurement of gas samples (such as the ABL 300, IL, or Corning blood gas machines). Note the Fick cardiac output formula from a previous lecture. Fick cardiac outputs are infrequently used mainly because of the inconvenience of collecting and analyzing exhaled gas concentrations. It's not as difficult to do as one might think but nonetheless Fick cardiac outputs are seldom used today. You may see mention of an estimated Fick cardiac output method where you just assume that oxygen consumption is normal by plucking a value off of a nomagram corrected for weight and height but in patients in whom a cardiac output determination is really needed, the oxygen consumption is seldom normal and these estimated cardiac output measurements can do more harm than good.

Dilution Methods

Dilution methods mathematically calculate (using calculus) the cardiac output based on how fast the flowing blood can dilute a marker substance introduced into the circulation normally via a pulmonary artery catheter. The marker must be distinguishable from the blood and must be able to be measured quickly and with a high degree of accuracy. Early dilution methods used dye solutions which were administered upstream and then drawn off in blood samples downstream from the infusion port where they could be analyzed for concentration. Cardiac output was inversely related to the downstream dye concentration. Dye dilution cardiac outputs are seldom used today outside of cardiac catheterization labs and even most of them use the more automated thermal dilution method. In thermal dilution, cold or room temperature water or D5W is used as the marker solution and distal concentration is determined by measuring the temperature downstream from the infusion port. Since water is non-toxic, multiple measurements can be made as often as needed and the downstream concentration (i.e. temperature) can be measured in situ without having to withdraw any blood from the circulation for analysis.

Errors

Cardiac output measurement is not precise using today's technology. For clinical use, we don't need 100% accuracy to 5 significant digits but to avoid big errors it is important to know the limitations of the measurement techniques. Fick cardiac output errors result from leaky gas collection apparatus, from inaccuracies in the measurement of inhaled and exhaled oxygen concentrations (these are particularly common when high levels of oxygen are used), an from errors in the calculations and/or measurements of blood oxygen contents (such as might be caused by using a bogus hemoglobin level or assuming the absence of carbon monoxide affecting oxygen saturation). Thermal dilution cardiac outputs are affected by the phase of respiration, particularly during mechanical ventilation and should thus always be measured at the same point in the respiratory cycle (normally end-expiratory) where the effect of breathing (either spontaneous or mechanical) is least. Small errors can result from using the wrong fluid (something other than D5W) as the injectate. Variations in the speed of cold water injection can result in altered measurements and devices to automatically inject the fluid are available to eliminate this source of variation. While there are lots of things that can result in cardiac output measurements not exactly equaling the true cardiac output, the most important concept here is to make the measurements reproducible and the errors consistent from one measurement to the next. It is the change in cardiac output, up or down, that allows the practitioner to determine the effects of therapy and disease and not the absolute value but to accurately detect changes, the output measurement errors must be consistent.

Pulmonary Circulation

Pulmonary & Systemic Circulation
By Donald R. Elton, MD, FCCP
Lexington Pulmonary and Critical Care

Introduction

We will concern ourselves with two main types of blood vessels: systemic and pulmonary. After birth, systemic vessels (arteries and veins) supply peripheral tissues, including the lung and other organs, with oxygen and nutrients and remove carbon dioxide. The pulmonary vessels are responsible for carrying deoxygenated blood to alveoli in the lungs where gas exchange with the atmosphere takes place. Note that the systemic arteries and pulmonary veins both carry oxygenated blood while the systemic veins and pulmonary arteries carry deoxygenated blood. Pulmonary tissues themselves are supplied partially by the relatively deoxygenated pulmonary arteries and partially by the systemic bronchial arteries. All systemic arteries arise from the left ventricle and aorta while all pulmonary arteries arise from the right ventricle. All systemic veins drain into the vena cava and right atrium while all pulmonary veins drain into the left atrium.

Systemic Circulation
The systemic circulation is a high pressure system. Pressures within systemic arteries in some disease states can be as high as 300 mm Hg though normal pressures are below 150 mm Hg. The Aorta, the body's largest systemic artery, arises from the left ventricle and provides branches that supply the entire body. The important coronary arteries arise from the aorta just above the aortic valve at the left ventricular outflow tract. Systemic blood pressure (and really all blood pressures) are determined at a basic level by flow and resistance, namely the cardiac output and systemic vascular resistance. Cardiac output, in turn, is determined by heart rate and stroke volume (the volume pumped by the left ventricle in a single contraction). The body regulates the systemic blood pressure both by regulating the cardiac output and the systemic vascular resistance. Cardiac output can be increased or decreased by changing contractility (which results in alterations in stroke volume) or by changing the heart rate. There are several feed-back mechanisms involving primarily the kidneys and brain that regulate blood pressure. The kidneys play an important role as a certain amount of renal perfusion is required for their proper function and the brain is involved as it has high oxygen demands with little energy storage capability. Systemic arteries subdivide into smaller and smaller vessels until they become capillaries capable of allowing only one red blood cell to pass at a time. These capillaries are where gas exchange occurs between the oxygenated systemic blood and peripheral tissues. Capillaries, then empty into systemic veins that lead back toward the vena cava and right atrium. Unlike arteries, the systemic veins frequently have valves that assist in preventing back flow away from the heart. These are necessary as veins are the low pressure end of the systemic circulation and in animals like the human or giraffe, the vertical distance the venous blood has to travel to reach the right heart requires more pressure to overcome than will normally exist in the veins. Muscular activity and negative intrathoracic pressure both help push and pull systemic venous blood back toward the heart but they are not continuous forces so the valves prevent blood that was recently pushed or pulled toward the heart from sliding back into the distal legs and arms. Typical systemic venous pressures are in the range of 0 to 10 mm Hg. While there are some variations in arterial anatomy (i.e. finding one artery replacing two) arteries are fairly constant from one individual to another. Veins are much more variable with only the larger ones being constant across individuals. Humans can regenerate veins or generate collateral veins with little difficulty while this is frequently not possible with arteries. One reason why it is serious to damage or obstruct an artery is that there may well be necrosis of the tissue supplied by the artery before any repair or replacement of vasculature can take place. Arteries that supply a tissue without any collateral (backup) artery are called end arteries. Coronary arteries are frequently end arteries. Loss or obstruction of an end artery results in tissue necrosis. Obstruction of a vein results in alternative veins becoming engorged and can result in edema in the affected area but these changes normally do not result in tissue necrosis.

Systemic Arterial Pressure Monitoring
Systemic arterial pressure measurement is done either with a manual or automatic blood pressure cuff (sphygmomanometer) or with an indwelling arterial catheter attached to a strain gauge pressure transducer via a fluid filled catheter. It is normally desirable to use the cuff method so long as pseudo-hypertension does not occur from hardening of arterial walls and so long as the blood pressure is not so low as to make the cuff readings unreliable (the cut-off varies from patient to patient but an arterial catheter would normally be used if the blood pressure is below 70 mm Hg). Arterial catheters are saved for life threatening conditions because they are frequently associated with complications such as bleeding (you can exsanguinate a patient in very short order if tubing disconnects inadvertently) and occlusion of the artery which can result in loss of the hand, fingers, or other distal structures. Clots can form on an arterial catheter which can embolize (break free) and the clots can travel downstream occluding distal arteries resulting in tissue necrosis. A less frequently recognized complication of arterial blood pressure monitoring that is particularly important when they are used in pediatric or neonatal patients but still valid in adults is the pressure that can be exposed to the circulation when flushing the catheter. Most monitoring systems use an IV bag pressurized to 300 mm Hg that is then regulated via a down-regulating device to allow a few ml's per hour of heparin-containing fluid to pass through the catheter to keep the catheter from clotting off. When blood is drawn through the catheter, there is a button or switch that will briefly allow the 300 mm Hg pressure to be applied directly to the catheter, bypassing the regulator. This can cause a rapid rise in arterial pressure depending on the diameter of the catheter in use and can dislodge clots from the catheter pushing them both downstream (their normal direction of travel) and upstream which could result in emboli being sent far from the artery where the catheter sits. This could even cause a stroke or blindness if clots were washed toward the head during a catheter flush. To improve the safety of arterial catheter usage, they should only be placed when absolutely necessary, they should not be used for routine blood draws (other than perhaps blood gasses), they should only be used in an intensive care unit setting where they can be closely watched, they should only be placed in vessels that have collaterals (i.e. not in femoral or brachial arteries), and they should be removed as soon as possible or as soon as any sign of a complication presents.

Systemic Arterial Oxygen Monitoring
In the absence of unusual shunts, arterial blood from any systemic artery will have the same oxygen content as blood from any other systemic artery.

Systemic Venous Pressure Monitoring
Venous pressure monitoring is only done in the large or central veins. This is to avoid local variations in limbs and to avoid the effect that valves have on the pressures. Normally this means a catheter must be placed into the vena cava or right atrium of the heart. Typical access sites would include the internal or external jugular veins or the subclavian veins or femoral veins. Because venous pressures are lower, they can be measured with a column or with a transducer like that used for arterial pressure monitoring. Columns are simpler and cheaper but are seldom used today because of the availability of transducers which are more accurate when used properly and offer continuous rather than intermittent monitoring. The central venous pressure (CVP) can be estimated by physical exam (see diagram above) by measuring the vertical distance between the top of the internal jugular venous pulsation in the neck and the sternal angle and adding 5 to the value to give you the CVP in units of cm H2O.

Systemic Venous Oxygen Monitoring
Systemic venous oxygen contents reflect the oxygenation of the tissues they drain. For this reason, if the practitioner needs to know the venous oxygen content, the blood must be drawn from an area where all of the systemic venous blood in the body has mixed. The only suitable location to find such blood is in the pulmonary artery. While mixing actually starts at the level of the vena cava, studies have shown that mixing is not complete until the pulmonary artery. Some Swan-Ganz pulmonary artery catheters have a built-in oximeter that can continuously monitor the oxygen saturation in the pulmonary artery.

Pulmonary Circulation
The pulmonary circulation is a low pressure system. Normal pressures are less than 30 mm Hg though in certain disease states the pulmonary artery pressure can be 70 mm Hg or more and may approach the systemic arterial pressure. There is very little blood flow through the pulmonary circulation before birth but after the first breath, the pulmonary vascular resistance falls rapidly allowing blood that previously passed through the patent ductus arteriosis and foramen ovale to pass through the lungs to pick up oxygen and lose carbon dioxide. Pressures are lower in the pulmonary circulation because the pulmonary vascular resistance is lower. In the absence of shunts, pulmonary and systemic blood flow are equal. The lower pressures are the reason why the right ventricle has a thinner wall than the muscular left ventricle which must deal with higher systemic blood pressures. The pulmonary vascular resistance is really never too low but it can be made too high by a variety of disease states.

Pulmonary Hypertension
Elevations in pulmonary vascular pressures result primarily from increases in vascular volume, increases in pulmonary vascular resistance, or from blockage or destruction of pulmonary vessels. Pulmonary vascular resistance can be increased by hypoxemia or acidemia or by some drug effects. The increase caused by hypoxemia or acidemia has a useful function as these are the conditions that exist when a part of the lung is underventilated. If you underventilate a part of the lung but do not change the pulmonary blood flow to that area then the blood going to the underventilated segment of lung will not be brought up to arterial oxygen levels and will be added to the oxygenated blood from the well ventilated segments as a shunt. By clamping down on vessels to underventilated areas, the lung prevents shunts from developing. Because the pulmonary circulation can handle large increases in cardiac output with minimal pressure elevations, shutting down blood flow to a segment or even lung usually does not result in significant pulmonary hypertension. However, if you globally reduce oxygenation or have acidosis, then all pulmonary vessels will clamp down resulting in an increase in the pulmonary vascular resistance and resultant pulmonary hypertension. This is an example of how a normal adaptive mechanism can cause pathology in some disease states. Obstruction of pulmonary vessels can occur as a result of pulmonary embolic disease or from destructive lung diseases such as emphysema or cystic fibrosis. Usually, 30-50% of the pulmonary circulation must be lost before significant pulmonary hypertension occurs. Pulmonary vascular pressure monitoring This will be covered during the lectures on the Swan-Ganz catheter.

Congenital Heart Disease
There are numerous congenital heart defects that result in alterations in normal anatomy and physiology of the pulmonary and systemic circulatory systems. Some defects result from abnormal shunts or openings between the two circulatory systems such as in atrial or ventricular septal defects or persistence of fetal circulation and other defects result in abnormal connections of normal vessels such as transposition of the great vessels or anomalous pulmonary venous return. Some of these defects cause right to left shunting (reduced pulmonary blood flow) and these defects result in severe hypoxia. Other defects cause left to right shunting (increased pulmonary blood flow) and these result in milder hypoxia resulting from pulmonary edema. It is important to have an idea of what sorts of defects can occur as many of these defects are barely symptomatic and can escape detection until adulthood and may be noted incidentally when a Swan-Ganz catheter is inserted for management of an unrelated condition but they can cause unexpected pressures and oxygen contents to be measured.